SurgiStat II Service Manual
Rear Panel
Figure 2-6.Layout of connectors
and controls on the rear panel
Equipotential Connector
Allows attaching a standard
grounding cable to chassis ground
for additional protection against
leakage current.
Volume Control
Controls the volume of the audible tones
produced during normal unit activation.
To increase volume, rotate the knob
clockwise.
Footswitch Receptacle
Accepts the E6008 or E6008B
monopolar footswitch. Use the
monopolar footswitch for both
monopolar and bipolar
activation. Use only a Valleylab
monopolar footswitch with a
SurgiStat II generator. Use of an
incompatible footswith may
cause unexpected output.
Serial Number Label
Specifies the unit model
number, serial number,
nominal line voltage,
frequency, current, and
power consumption.
Power Cable Receptacle
and Fuse Holder
Connects a hospital grade
power cord to supply AC
mains power to the unit.