SurgiStat II Service Manual
3-5
Low Frequency (50-60 Hz) Leakage Current
High Frequency (RF) Leakage Current
Standards and IEC Classifications
The SurgiStat II generator meets all pertinent clauses of the IEC 60601-1 second
edition and IEC 60601-2-2 third edition.
Class I Equipment (IEC 60601-1)
Accessible conductive parts cannot become live in the event of a basic insulation
failure because of the way in which they are connected to the protective earth
conductor.
Type CF Equipment (IEC 60601-1)/Defibrillator Proof
Enclosure source
current, ground open
< 300 µA
90V-110V mains voltage
110V-120V
< 500 µA
220V-240V mains voltage
Source current, patient
leads, all outputs
Normal polarity, intact ground: < 10 µA
Normal polarity, ground open: < 50 µA
Reverse polarity, ground open: < 50 µA
Sink current at high
line, all inputs
< 50 µA
Bipolar RF leakage
current
< 39 mA
rms
Monopolar RF leakage
current
< 150 mA
rms
The SurgiStat II generator provides a high degree of protection
against electric shock, particularly regarding allowable leakage
currents. It is type CF isolated (floating) output and may be used
for procedures involving the heart.