T
e
chnical S
pecificatio
ns
SurgiStat II Service Manual
3-9
Guidance and manufacturer's declaration - electromagnetic immunity
The SurgiStat II is intended for use in the electromagnetic environment specified below. The customer or the user of
the SurgiStat II should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Conducted RF IEC
61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150KHz to 80MHz
3 V/m
80MHz to 2.5GHz
3 V
3 V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of the
SurgiStat II, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d=1.2
√
P
d=1.2
√
P 80MHz to 800MHz
d=2.3
√
P 800MHz to 2.5GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1
At a 80MHz and 800MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the SurgiStat II is used exceeds the
applicable RF compliance level above, the SurgiStat II should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the SurgiStat II.
b.
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
Summary of Contents for SurgiStat II
Page 1: ...Service Manual SurgiStat II Electrosurgical Generator ...
Page 24: ...2 12 SurgiStat II Service Manual ...
Page 46: ...3 22 SurgiStat II Service Manual Figure 3 20 Bipolar mode waveform ...
Page 60: ...5 6 SurgiStat II Service Manual ...
Page 64: ...6 4 SurgiStat II Service Manual ...
Page 74: ...7 10 SurgiStat II Service Manual ...