T
roubleshooting
SurgiStat II Service Manual
7-5
E5
Pulse width
error
Internal pulse width measurement
exceeds setting requirements
1.
Turn unit off.
2.
If using a metal patient return electrode plate, verify
that the cable connections are secure and replace
cables with loose connections. Turn unit on.
3.
If error code occurred directly after changing mode
of operation, turn on the unit and delay activation of
the accessories after changing mode of operation.
4.
If alarm number reappears upon restarting the unit,
record the number and call the Valleylab Service
Center.
E6
Delta check
error
Internal pulse width measurement
conflicts with calibrated value
1.
Turn unit off.
2.
If error code occurred directly after changing mode
of operation, turn on the unit and delay activation of
the accessories after changing mode of operation.
3.
If error code occurred directly after changing the
power setting during activation, turn on the unit and
limit the change of power to approximately three
watts per second.
4.
If alarm number reappears upon restarting the unit,
record the number and call the Valleylab Service
Center.
E7
Temperature
error
Internal temperature of the unit
exceeded limit
1.
Turn the unit off.
2.
Allow the unit to cool for 20 minutes.
3.
Restart the unit.
E8
Connector
sensing error
caused by a
sensed open circuit
The 24-pin ribbon cable connected
between the main board and the
display board is either disconnected
or damaged.
Contact your Valleylab Service Center.
Error Code
Description
Recommended Action
Summary of Contents for SurgiStat II
Page 1: ...Service Manual SurgiStat II Electrosurgical Generator ...
Page 24: ...2 12 SurgiStat II Service Manual ...
Page 46: ...3 22 SurgiStat II Service Manual Figure 3 20 Bipolar mode waveform ...
Page 60: ...5 6 SurgiStat II Service Manual ...
Page 64: ...6 4 SurgiStat II Service Manual ...
Page 74: ...7 10 SurgiStat II Service Manual ...