background image

Setup

9

Once the phase contrast condenser assembly has been centered, the phase contrast 

annulus rings must be centered for the objectives.  Phase contrast is a system which 

involves  a  series  of  light  baffling  annular  rings.    Proper  alignment  of  these  rings  is 

absolutely necessary to achieve phase contrast.

Aligning the Phase Contrast Annulus Rings

Begin by turning on the substage illuminator with the power switch.  
Set the objectives so they are in the approximate position for actual use.  This is best 

achieved by placing a slide on the stage, rotating the 100X objective into position, 

and raising the stage until the tip of the 100X objective is just above the slide (almost 

touching).  
Rotate the nosepiece until the 10X objective is in the light path.

Rotate the annuli turret in the phase contrast condenser assembly until the “10” is 

seen in the viewing window.  
Remove an eyepiece from one of the eyetubes and replace with the phase contrast 

centering telescope.

Looking through the phase contrast centering telescope, turn the uppermost piece 

of the phase contrast centering telescope until the image is in focus.
The image seen through the phase contrast centering telescope should resemble 

rings superimposed on one another (see Figure 7).  What is actually being viewed 

are the phase rings.  Holding a white sheet of paper on the stage and sliding back 

and forth between the objective and stage will make this easier to see.
Turn the phase annuli centering knobs, which extend from the phase contrast 

condenser assembly, until the two rings of light are centered upon one another (see Figure 8).

Once the phase rings are centered, remove the phase contrast centering telescope and replace with the eyepiece.
The phase rings are now centered for the other remaining objectives.  This process shouldn’t need to be repeated 

for each objective setting, although it is advised to perform off and on checks with the phase contrast centering 

telescope to confirm that the phase rings are still centered.

NOTE:  Brightfield work can be achieved on models with a phase contrast condenser.  The “0” setting on 

the phase annulus turret is used for this purpose.

Verifying that the illuminator is centered will allow maximum light to reach the specimen.  This adjustment is usually 

not required as our microscopes are shipped with the illuminator pre-centered.  If for some reason the illuminator 

should need to be centered again follow the instructions below.  

Centering the Illuminator

Remove the substage collector by loosening the set screw on the left hand side.

The bulb filament should be centered in the opening.  If it is not, loosen the bulb centering mechanism on 

the bottom of the microscope.

Reposition the bulb so that the filament is at the center of the opening.  Make sure that the reflector beneath 

the bulb is aligned with the light bulb access door so that the access door can still be opened.
Tighten the bulb centering mechanism.
Reinstall the substage collector.

3

4

1

1

3
4

5

6

7

8

9

10

5

All Models

Phase Contrast Models

If you purchased a brightfield model this completes your setup.  Skip ahead to page 10, “Using your 

1200CM Series Microscope”.  If you purchased a phase contrast model, continue below.

Figure 7

Figure 8

Summary of Contents for 1200CM Series

Page 1: ...Covering Models 1210CM 1211CM 1220CM 1221CM 1222CM 1223CM 1230CM 1231CM 1232CM 1233CM Operation Manual 1200CM Series...

Page 2: ...25 N A Abbe condenser Phase contrast models come with a 1 25 N A Zernike condenser with phase annulus rings for 10X 25X 40X and 100X also has a 0 setting for brightfield work All condensers are mount...

Page 3: ...0X Plan Phase Objective 1 ea 20X Plan Phase Objective 1 ea 40X Plan Phase Objective 1 ea 100X Plan Phase Objective 1 ea Dispersion Filter 1 ea Blue Filter 1 ea Green Filter 1 ea Model 1230CM Trinocula...

Page 4: ...xels for 1210 and 1220 series 1200 CDPC 3 USB DIgital Camera System 3 Megapixels for 1230 series Other Accessories Part Number Description Call for Part Numbers Eyepieces 10X 16X 20X Call for Part Num...

Page 5: ...tering Knobs Brightfield only Aperature Iris Diaphragm Adjustment Lever Fine Focus Control Substage Collector Base Stand Mechanical Stage Slide Holder Slide Holder Knobs Focusing Tension Control Knob...

Page 6: ...Fuse Holder Parts Accessories Phase Annuli Centering Knobs Phase Contrast Condenser Iris Diaphragm Adjustment Light Bulb Centering Mechanism Substage Collector Set Screw 6 Light Bulb Access Door Lock...

Page 7: ...eyetubes see figure 2 After removing the objectives from their storage containers individually install each one into the nosepiece by threading them clockwise into the threaded holes of the nosepiece...

Page 8: ...er will reduce the field so that a small white hexagon is visible within a black field see Figure 6A Focusing of the hexagon is performed by turning the coarse fine focus controls This white hexagon i...

Page 9: ...nser assembly until the two rings of light are centered upon one another see Figure 8 Once the phase rings are centered remove the phase contrast centering telescope and replace with the eyepiece The...

Page 10: ...to induce contrast into a specimen not to adjust light intensity The mechanical stage X Y controls located underneath the right hand side of the stage see figure 9 provide easy and accurate positionin...

Page 11: ...icroscope Interpupillary and Diopter Adjustment 1 2 3 Interpupillary adjustment the distance between eyepieces is made through a folding action The Seidentopf design allows for a folding adjustment wh...

Page 12: ...a adapter assembly until it is firmly secured with the camera port set screw Tighten the camera port set screw Pull the prism slider completely out to divert the image to the camera port The prism sli...

Page 13: ...Models The 100X objective which comes with this microscope must be used with immersion oil in order to maintain image quality After use the objective tip needs to be wiped clean so that no oil residu...

Page 14: ...lose the light bulb access door and screw the knob back in to lock Return the microscope to normal operating position If necessary recenter the illuminator following the instructions on page 9 1 2 7 6...

Page 15: ...r ordering information The same care instructions apply to all optical parts on this VanGuard Microscope including the substage collector and the condenser lenses All other parts can be cleaned with a...

Page 16: ...can t get enough light Step Possible Cause How to Check Result Solution For More Info 1 Variable lighting control VLC is turned down too far Visually inspect VLC turned up Go to next step VLC turned...

Page 17: ...blockage Contact dealer or VanGuard Microscopes Page 17 I can t focus on the specimen Step Possible Cause How to Check Result Solution For More Info 1 Up stop set too low Visually inspect Set too low...

Page 18: ...25 NA 8 0mm WD 1 8mm FOV Achromatic or Plan Achromatic 20X 0 40 NA 0 50mm WD 0 9mm FOV 40X 0 65 NA 0 40mm WD 0 45mm FOV 100X oil 1 25 NA 0 25mm WD 0 18mm FOV Stage Dimensions 125mm x 135mm Stage Motio...

Page 19: ...19 Notes...

Page 20: ...Printed in the USA No VGMNL070319 1200CM...

Reviews: