Setup
Note: This completes the setup for the 1486FLi. The next 5 pages will explain how to use and make adjustments
to the microscope.
12
1.
Begin by turning on the
substage illuminator
with the
power switch
located on the
lower left side of the instrument. Set the
objectives
so they are in the approximate
position for actual use. This is best achieved by placing a slide on the
stage
,
rotating the 100X
objective
into position, then raising the stage (via the
coarse/
fine focus control
knobs) until the tip of the 100X
objective
is just above the slide
(almost touching).
2.
Rotate the
nosepiece
until the 10X
objective
is in the light path, then rotate the
phase annulus turret
(the dial on the front of the condenser) in the
phase contrast
condenser assembly
until the “10” is seen in the viewing window (see figure 16).
3.
Remove an
eyepiece
from one of the
eyetubes
and replace with the
phase contrast
centering telescope
(see figure 17).
4.
After loosening the set screw on the
phase contrast centering telescope
, look
through the
phase contrast centering telescope
and
pull out the uppermost piece
of the
phase contrast centering telescope
until the image is in focus. Tighten the
set screw.
5.
The image seen through the
phase contrast centering telescope
should resemble
rings superimposed on one another (see Figure 18A). What is actually being
viewed are the phase rings.
6.
Turn the
condenser centering knobs
, which extend from the front of the condenser
mount, until the two rings of light are centered upon one another (see Figure 18B).
Aligning the Phase Contrast Annulus Rings
Phase contrast is a system which utilizes a series of light baffling annulus rings. Proper
alignment of these rings is absolutely necessary to achieve optimal phase contrast.
Figure 7B
Figure 7A
1486FLi Model
Figure 17
Figure 16
7.
Once the phase rings are centered, remove the
phase contrast centering telescope
and replace with the
eyepiece
.
The phase rings are now centered for the other remaining
objectives
unless one of the factory phase objectives has been
replaced. In this instance you will need to align just that condenser setting for your new objective using the centering knobs
located at the rear of the condenser (not the knobs at the front as used in step 6 above).
NOTE: Brightfield work can be achieved by setting the condenser at the “0” marking. Likewise, the “DF” setting is
for Darkfield work.