Care and Maintenance

The Compact Turbimeter has been tested for IP67 compliance so accidental 

immersion will not damage the internal components. Sample cuvette should be 

clean and dry prior to insertion into the sample holder. Lint free cloths are provided 

for this purpose.

Silicon oil is also present to use on scratched cuvette surfaces its use should not be 

necessary unless turbidity readings are consistently higher than expected. If required 

place two or three drops onto the vial and smear across the glass using the lint free 

cloth.

User calibration is advised on a periodic basis, using the supplied SDVB standards. 

Calibration frequency is at the user discretion and can be deduced by reading the 

standards to determine any offset that may have occurred due to fouling.

Subjecting the standards to prolonged exposure to low temperatures will reduce the 

NTU value. The use of a WP48 Heat Pad is recommenced in these circumstances to 

prevent standard degradation. Standards have a 12 month stability and should be 

replaced after this date. No prescribed maintenance intervals are possible due to 

the high dependence on frequency of use, types of samples and conditions and 

user protocol. Service and maintenance is available via authorised service centres 

or direct from Palintest Ltd. 

Troubleshooting

The Compact Turbimeter has internal diagnostics to advise on some issues as follows:

Error 100 – Measurement Off Scale

The reading is above 1050 NTU. Dilute using the provided PT 512 Dilution Tube and 

clean water and re-measure.

Error 107- Blanking Error

The sample cuvette cap is not attached allowing ambient light into the 

QuadoptiX™ chamber. Replace the cap and read again.

Error 110 – Battery Low

The available power is insufficient to support reliable analysis. Replace the batteries.

Other more general issues that arise may be as follows :

Summary of Contents for CT12

Page 1: ...x 01428 656 808 www vanwalt com Compact Turbimeter 1 VAN WALT Monitoring your needs CONTENTS INTRODUCTION KIT CONTENTS QUADOPTIX TECHNOLOGY START UP BATTERY LIFE SYSTEM MENU READING MENU CALIBRATION M...

Page 2: ...TENTS Compact Turbimeter Instrument CT12 Model 4 x Sample Cuvettes 4 x SDVB Calibration Standards Silicone Oil 10 ml 2 x Packs of Lint Free Cloths Sample Dilution Tube Cleaning Brush 2 x Heat Pads WP4...

Page 3: ...ON After initialization the Mode screen will appear and display three options 1 Reading Select the mode to measure turbidity or suspended solids 2 Calibration Choose either SDVB Total Suspended Solid...

Page 4: ...r data including the last mode used for reading Replacing the AA batteries will not lead to settings or calibration data being lost If battery power is insufficient for effective operation the Compact...

Page 5: ...by pressing Done briefly Select OK to accept or Edit to change the entry Multiple Operator IDs can be stored and recalled from memory as required Sample ID Optionally set the Sample ID using the same...

Page 6: ...event stray light from adversely affecting readings Microscopic scratches will refract light and can lead to higher than expected values To prevent scratches having an impact apply a thin film of sili...

Page 7: ...ng mode select Normal and press OK Insert the sample with the orientation mark facing forward and press Read A beep will sound at the start and end of measurement Result is displayed in and stored in...

Page 8: ...wn arrows The calibration data will be recalled automatically Insert the sample with the orientation mark facing forward and press Read A beep will occur at the start and end of each measurement cycle...

Page 9: ...can be used 2 Factor Using data manipulated externally a slope and intercept for the straight line relationship can be entered directly Calibrating the Compact Turbimeter SDVB Select Calibration and p...

Page 10: ...ady been created use the Up Down arrows to highlight and press OK to select If a new Sample ID is required move the highlight to a blank area and press New Create the new Sample ID using the Up Down b...

Page 11: ...mine any offset that may have occurred due to fouling Subjecting the standards to prolonged exposure to low temperatures will reduce the NTU value The use of a WP48 Heat Pad is recommenced in these ci...

Page 12: ...materials Readings are unstable variable Turbidity is a measurement of scattered light with the source of scattering being microscopic and larger solids present in the sample Solids can be mobile or...

Reviews: