Operating Instructions

Van Walt Ltd | Prestwick Lane | Grayswood | Haslemere | Surrey | GU27 2DU | Tel. 01428 661 660 | Fax. 01428 656 808 | www.vanwalt.com

Compact Turbimeter

8

VAN WALT

Monitoring your needs

 

Total Suspended Solids – When using the Compact Turbimeter in Total Suspended 

Solids Reading mode, a calibration is required to set the relationship between the 

measured turbidity values and the true suspended solids value, determined by 

gravimetric methods. This calibration can either be an assigned series of values 

from the results log or a simple algorithm with factors provided by the user. This 

relationship is typically a straight line and hence the factors consist of a slope and 

intercept. The Compact Turbimeter has two methods of generating the data for 

relating the measurement of turbidity to suspended solids :

1.  Correlation – Using the stored turbidity data from the instrument log and relating 

it to manually entered Suspended Solids data derived from elsewhere. The 

minimum number of correlation points required is one as the zero point is used 

by default. A maximum number of 50 data points can be used.

2.  Factor – Using data manipulated externally, a slope and intercept for the 

straight line relationship can be entered directly.  

Calibrating the Compact Turbimeter

SDVB

Select Calibration and press ‘OK’. Choose SDVB and press ‘OK’. Select the 0.10 

NTU standard and insert into the sample chamber with the orientation mark facing 

forward. Use the Up/Down buttons to adjust the value on screen to that on the 

standard and press ‘Read’. Repeat the process for the 20 NTU, 100 NTU and 800 

NTU standards. On completion of the process the Compact Turbimeter screen will 

show ‘Successful’. Press ‘Back’ to leave the calibration process and return to the 

Calibration menu. 

To cease calibration during the data capture phase press ‘Cancel’. To cease the 

calibration process at any step press ‘Back’ (calibration data will be discarded).

The time, date and Operator ID are stored in the calibration log for GLP compliance.

Total Suspended Solids – Correlation

Select Calibration and press ‘OK’. Select Total Suspended Solids and press ‘OK’.

If the Sample ID has already been created use the Up/Down arrows to highlight and 

press ‘OK’ to select. If a new Sample ID is required, move the highlight to a blank 

area and press ‘New’. 

Create the new Sample ID using the Up/Down buttons to select characters as 

describe in the System section.

Summary of Contents for CT12

Page 1: ...x 01428 656 808 www vanwalt com Compact Turbimeter 1 VAN WALT Monitoring your needs CONTENTS INTRODUCTION KIT CONTENTS QUADOPTIX TECHNOLOGY START UP BATTERY LIFE SYSTEM MENU READING MENU CALIBRATION M...

Page 2: ...TENTS Compact Turbimeter Instrument CT12 Model 4 x Sample Cuvettes 4 x SDVB Calibration Standards Silicone Oil 10 ml 2 x Packs of Lint Free Cloths Sample Dilution Tube Cleaning Brush 2 x Heat Pads WP4...

Page 3: ...ON After initialization the Mode screen will appear and display three options 1 Reading Select the mode to measure turbidity or suspended solids 2 Calibration Choose either SDVB Total Suspended Solid...

Page 4: ...r data including the last mode used for reading Replacing the AA batteries will not lead to settings or calibration data being lost If battery power is insufficient for effective operation the Compact...

Page 5: ...by pressing Done briefly Select OK to accept or Edit to change the entry Multiple Operator IDs can be stored and recalled from memory as required Sample ID Optionally set the Sample ID using the same...

Page 6: ...event stray light from adversely affecting readings Microscopic scratches will refract light and can lead to higher than expected values To prevent scratches having an impact apply a thin film of sili...

Page 7: ...ng mode select Normal and press OK Insert the sample with the orientation mark facing forward and press Read A beep will sound at the start and end of measurement Result is displayed in and stored in...

Page 8: ...wn arrows The calibration data will be recalled automatically Insert the sample with the orientation mark facing forward and press Read A beep will occur at the start and end of each measurement cycle...

Page 9: ...can be used 2 Factor Using data manipulated externally a slope and intercept for the straight line relationship can be entered directly Calibrating the Compact Turbimeter SDVB Select Calibration and p...

Page 10: ...ady been created use the Up Down arrows to highlight and press OK to select If a new Sample ID is required move the highlight to a blank area and press New Create the new Sample ID using the Up Down b...

Page 11: ...mine any offset that may have occurred due to fouling Subjecting the standards to prolonged exposure to low temperatures will reduce the NTU value The use of a WP48 Heat Pad is recommenced in these ci...

Page 12: ...materials Readings are unstable variable Turbidity is a measurement of scattered light with the source of scattering being microscopic and larger solids present in the sample Solids can be mobile or...

Reviews: