background image

Setup

Setup

Setup

Setup

Setup

10

19

1.

Focusing adjustment is achieved by turning the coarse/fine focus controls
located on both sides of the microscope. The large knob is used for coarse
adjustment, the smaller knob for fine adjustment (figure 48). The coaxial
arrangement allows for easy, precise adjustment without stage drift.

Focusing and Mechanical Stage Mechanisms

Microscope Use

Microscope Use

Microscope Use

Microscope Use

Microscope Use

Figure 49

Figure 50

Figure 51

3.

The mechanical stage X-Y controls, located underneath the right-hand side
of the stage (figure 49), provide easy and accurate positioning of the
specimen. One complete turn of the longitudinal (Y-axis) control (lower half
of the stage controls) will move the specim

en 20mm left or ri

ght. One

complete turn of the transverse (X-axis) control (upper half of the stage
controls) will move the s

pecimen 33mm front or ba

ck.

2.

Turning the coarse/fine focus control raises and lowers the stage vertically.
One complete turn of the fine focusing knob ra

ises or lowers the stage

0.2mm; the smallest graduation refers to 2 microns of v

ertical movement.

One half turn of the coarse focusing knob raises 

or lowers the stage 20mm.

To ensure long life, turn the focusing knobs slowly and uniformly. Care should
always be taken when using the focus controls to prevent the objectives
from making contact with the stage or specimen.

Abbe Condenser Swing-In Lens

4.

The transmitted illumination Abbe condenser features a swing-in lens located
at the bottom of the condenser (figure 50). This lens should be swung into
the light path when using the 5X objective only. This is necessary to obtain
the correct Numerical Aperture required for the low powered objective. In
addition you will need to fully open the aperture iris in order to prevent
vignetting and leave the condenser height the same as when originally setup
using the 10X objective.

5.

To control contrast in your specimen image while operating with the 5X
objective use the field diaphragm (figure 51) instead of the aperture diaphragm
that is typically used.

7.

Insert the round electrical plug from the lamp housing into the jack located
on the rear of the stand (figure 6). Tighten the outer collar to prevent the
cable from coming loose.

Head and Reflected Illuminator Assembly

 

(cont.)

Objective Installation

10.

Adjust the coarse focus knobs so that the stage is completely lowered.

11.

Remove the objectives from their packing canisters and thread each one
into the nosepiece holes ensuring they are fully seated (figure 9). Typically
they are installed in numerical order from lowest to highest magnification.

Figure 6

Figure 7

Figure 8

Figure 9

Figure 10

8.

Insert the dovetail flange located on the bottom of the head into the hole at
the top of the reflected illuminator. The eyetubes should be facing forward.

While still holding the head

 tighten the head setscrew with the 2mm hex

wrench (figure 7).

9.

Remove the two packing caps from the eyetube openings and set aside
with the other packaging as they will not be needed. Slide the eyepieces into
the eyetubes and ensure they are fully seated. If you wish you may attach
the eyecups to the ends of the eyepieces, though this step is optional and
eyeglass wearers especially may choose not to install the them (figure 8).

Stage Plate Installation

12.

With the stage completely lowered place the glass stage plate into the
rectangular recess on top of the stage (figure 10). Clean the plate first if
necessary.

collar

eyepieces

eyecups

objectives

stage plate

Figure 48

right side

focus

controls

left side

focus

controls

y-axis

x-axis

swing-out lens

field diaphragm

Summary of Contents for 1442MMi

Page 1: ...Printed in the USA VGMNL062512 1400MMi Operation Manual Model 1442MMi www veegee com 800 423 8842 sales veegee com support veegee com...

Page 2: ...croscope system Modifying the product in any manner for use not originally intended shall automatically void the manufacturer s warranty In no event shall VEE GEE Scientific Inc be held liable for any...

Page 3: ...surfaces are made Shock Hazard failure to follow instructions could lead to electrical shock Safety Precautions Before replacing the lamp or fuses always turn off the power switch and disconnect the p...

Page 4: ...ymptom Image through eyepieces is too dim Step Possible Cause HowToTest TestResult Solution For More Info 1 2 Variable lighting control VLC is turned down too far Visually inspect VLC is turned up VLC...

Page 5: ...turned down Go to next step Turn VLC up Page16 Microscope not getting power Inspect outlet Good outlet Outlet not good Go to next step Plug power cord into working outlet Blown fuse Visually inspect...

Page 6: ...an to prevent contaminate ingress is essential to the longevity of the objectives All other parts can be cleaned with a paper towel and mild detergent It s not recommended to use rubbing alcohol for c...

Page 7: ...eplacement Fuse 3A 250V Cat No 1400 FS5 If the microscope is plugged in but the illuminator is not turning on the fuse could be blown The instructions below detail how to check for a blown fuse You ca...

Page 8: ...resistance do not force 4 Close the compartment door and follow the instructions in the Reflected Illumination Setup section to properly center the lamp Replacing the Reflected Illuminator Lamp Replac...

Page 9: ...lyzer knob located on the right side of the reflected illuminator must be pushed inward to engage this function figure 55 1 Carefully remove all components from foam packaging and plastic bags Use the...

Page 10: ...ve In addition you will need to fully open the aperture iris in order to prevent vignetting and leave the condenser height the same as when originally setup using the 10X objective 5 To control contra...

Page 11: ...er installed but should you need to reinstall at any point follow the directions below 13 Rotate the nosepiece so that either the 5X objective or an empty slot is in the light path 14 Using the coarse...

Page 12: ...making adjustments to the illuminator housing or lamp fixture 3 Ensure that the illumination selector knob is pushed all the way inward R figure 17 Figure 15 Figure 17 Figure 18 Figure 19 4 Rotate the...

Page 13: ...b adjusts for latitudinal X axis and the lower for longitudinal Y axis movement figure 21 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 10 Rotate each of these in succession until the field of vie...

Page 14: ...and should be checked and readjusted before each use Focus Adjustment 18 Focusing on your specimen is achieved through rotating the coarse and or fine focusing knobs These are located on either side o...

Reviews: