background image

Setup

Setup

Setup

Setup

Setup

11

18

Setup

Setup

Setup

Setup

Setup

36.

Open the field diaphragm until the iris is just barely no longer visible
(figure 45).

37.

To eliminate stray light glare close the aperture iris until you have achieved
acceptable contrast in your specimen.

Figure 45

Figure 46

Figure 47

Kohler Illumination Setup

 

(continued)

Trinocular Camera Port Setup

Installation and setup of your camera system, if applicable, will vary depending
on the camera type and adapters that are utilized. Please refer to the instructions
that came with your camera adapter system for specific details.

38.

In general though, most systems will require the removal of the trinocular
port dust cap which is held in place by a set screw located at the rear of the
port (figure 46).

39.

Use the prism selector knob, located on the left side of the microscope
head (figure 47), to divert the image from the eyepieces to the camera. The
system uses a 100/0 split sliding prism to divert 100% of the light to the
camera port when selected. This provides the brightest image possible to
the camera which can be critical for certain photomicrographic applications.

This concludes the setup of your VanGuard 1442MMi industrial microscope. At
this point your microscope is ready for use. Continue for instructions on the
proper use of the instruments main components.

Abbe Condenser Installation

Your microscope will have come with the brightfield condenser installed but should
you need to reinstall at any point follow the directions below.

13.

Rotate the nosepiece so that either the 5X objective or an empty slot is in
the light path.

14.

Using the coarse focus knobs raise the stage completely. Be careful to not
crash the stage into the objectives. If the stage won’t raise fully check to see
if the stage limit is set too low. Refer to the Stage Limit instructions in the
Setup section of this manual for more information.

15.

Adjust the substage control so that the substage assembly is in the full
downward position (figure 11).

16.

Slip the brightfield condenser into the collar from below. The narrow end of
the condenser should be facing upward. With the iris control facing forward,
and centered properly, tighten the condenser set knob (figure 12).

This concludes the basic assembly of your VanGuard 1442MMi industrial mi-
croscope. Continue on to setup the optical alignment, illumination, and ergo-
nomic features for proper use.

Power Cord Installation

17.

Check to see that the main power switch is in the off (“0”) position
(figure 13).

Figure 11

Figure 12

Figure 13

Figure 14

18.

Connect the main power cord to the jack located at the rear of the
microscope stand (figure 14) and plug the other end into a grounded power
source.

19.

The microscope electrical system will accept either a 110V/60Hz or a 220V/
50Hz AC voltage source. If you ordered a 110V microscope it will have come
with a 3-Pin North American (NEMA 5) style power cord. A grounded Schuko
(CEE 7/7) style cord is included with the 220V microscopes.

substage

control

setscrew

power

switch

power

jack

setscrew

dust cap

selector

knob

Summary of Contents for 1442MMi

Page 1: ...Printed in the USA VGMNL062512 1400MMi Operation Manual Model 1442MMi www veegee com 800 423 8842 sales veegee com support veegee com...

Page 2: ...croscope system Modifying the product in any manner for use not originally intended shall automatically void the manufacturer s warranty In no event shall VEE GEE Scientific Inc be held liable for any...

Page 3: ...surfaces are made Shock Hazard failure to follow instructions could lead to electrical shock Safety Precautions Before replacing the lamp or fuses always turn off the power switch and disconnect the p...

Page 4: ...ymptom Image through eyepieces is too dim Step Possible Cause HowToTest TestResult Solution For More Info 1 2 Variable lighting control VLC is turned down too far Visually inspect VLC is turned up VLC...

Page 5: ...turned down Go to next step Turn VLC up Page16 Microscope not getting power Inspect outlet Good outlet Outlet not good Go to next step Plug power cord into working outlet Blown fuse Visually inspect...

Page 6: ...an to prevent contaminate ingress is essential to the longevity of the objectives All other parts can be cleaned with a paper towel and mild detergent It s not recommended to use rubbing alcohol for c...

Page 7: ...eplacement Fuse 3A 250V Cat No 1400 FS5 If the microscope is plugged in but the illuminator is not turning on the fuse could be blown The instructions below detail how to check for a blown fuse You ca...

Page 8: ...resistance do not force 4 Close the compartment door and follow the instructions in the Reflected Illumination Setup section to properly center the lamp Replacing the Reflected Illuminator Lamp Replac...

Page 9: ...lyzer knob located on the right side of the reflected illuminator must be pushed inward to engage this function figure 55 1 Carefully remove all components from foam packaging and plastic bags Use the...

Page 10: ...ve In addition you will need to fully open the aperture iris in order to prevent vignetting and leave the condenser height the same as when originally setup using the 10X objective 5 To control contra...

Page 11: ...er installed but should you need to reinstall at any point follow the directions below 13 Rotate the nosepiece so that either the 5X objective or an empty slot is in the light path 14 Using the coarse...

Page 12: ...making adjustments to the illuminator housing or lamp fixture 3 Ensure that the illumination selector knob is pushed all the way inward R figure 17 Figure 15 Figure 17 Figure 18 Figure 19 4 Rotate the...

Page 13: ...b adjusts for latitudinal X axis and the lower for longitudinal Y axis movement figure 21 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 10 Rotate each of these in succession until the field of vie...

Page 14: ...and should be checked and readjusted before each use Focus Adjustment 18 Focusing on your specimen is achieved through rotating the coarse and or fine focusing knobs These are located on either side o...

Reviews: