Lean the microscope back to expose the lamp adjustment control located
on the bottom of the microscope stand taking care not to let the glass stage
plate and/or speciments fall. Loosen the transmitted lamp adjustment knob
located on the lamp compartment door by turning counter clockwise
(figure 40).

Figure 40

Figure 41

Figure 43

Figure 44

17

Setup

Setup

Setup

Setup

Setup

Kohler Illumination Setup

When using the transmitted illuminator proper condenser alignment, iris settings,
and focusing are necessary to obtain true Kohler illumination and achieve the
highest quality image of your specimen.

33.

Open the transmitted illuminator aperture and field diaphragms fully (figure
42), rotate the objective turret so that the 10X objective is in the optical path,
and focus on a specimen.

Transmitted Illumination Setup

 

(continued)

34.

While viewing through the microscope and slowly closing the field diaphragm,
adjust the substage condenser vertical control until the edge of the field iris
is in focus (figure 43).

35.

The two control knobs can be turned in unison or independantly to position
the condenser in alignment with the field diaphragm and light path (figure
44).

Continued on next page...

Reflected Illumination Setup

1.

Set the main power switch to the ON position (“|”) and the illuminator selector
switch to the UP position (“||”) (figure 15).

The reflected illuminator housing will get hot if left on for
extended periods of time. Be careful when making
adjustments to the illuminator housing or lamp fixture.

!

3.

Ensure that the illumination selector knob is pushed all the way inward (“R”)
(figure 17).

Figure 15

Figure 17

Figure 18

Figure 19

4.

Rotate the Aperture and Field diaphragm dials to the right (figure 18) to fully
open the iris’s.

5.

Rotate the color filter wheel until the blank position is in the light path
(figure 18).

6.

Pull the Polarizer filter slider out until you feel a click. Pull the Analyzer selector
knob fully out (figure 18).

7.

Rotate the 5X objective into the light path. You may have to lower the stage
in order not to crash the higher magnification objectives into the stage.

8.

Look through the eyepieces and you should see a white field of view. You
may have to adjust the illuminator intensity dial in case the light is too dim or
too bright. If the brightest portion of the field of view isn’t centered
(figure 19) you will need to perform the following instructions.

Continued on next page...

selector

switch

power

switch

selector

knob

analyzer

selector

iris dials

filter

wheel

filter slider

31.

Move the knob in latitudinal and/or longitudinal directions (figure 40) until the
field of view as seen through the eyepieces is at it’s brightest overall and the
brightest portion is well centered in the field of view (figure 41). This step
may be aided by placing a bright white piece of paper on the stage directly
under the objective and the light field viewed through that.

32.

 Retighten the lamp adjustment knob by turning clockwise.

Figure 42

field iris

aperture

centering

controls

2.

Adjust the VLC (variable lighting control) (figure 16) counter-clockwise to
the maximum position.

Figure 16

VLC

adjustment

knob

Summary of Contents for 1442MMi

Page 1: ...Printed in the USA VGMNL062512 1400MMi Operation Manual Model 1442MMi www veegee com 800 423 8842 sales veegee com support veegee com...

Page 2: ...croscope system Modifying the product in any manner for use not originally intended shall automatically void the manufacturer s warranty In no event shall VEE GEE Scientific Inc be held liable for any...

Page 3: ...surfaces are made Shock Hazard failure to follow instructions could lead to electrical shock Safety Precautions Before replacing the lamp or fuses always turn off the power switch and disconnect the p...

Page 4: ...ymptom Image through eyepieces is too dim Step Possible Cause HowToTest TestResult Solution For More Info 1 2 Variable lighting control VLC is turned down too far Visually inspect VLC is turned up VLC...

Page 5: ...turned down Go to next step Turn VLC up Page16 Microscope not getting power Inspect outlet Good outlet Outlet not good Go to next step Plug power cord into working outlet Blown fuse Visually inspect...

Page 6: ...an to prevent contaminate ingress is essential to the longevity of the objectives All other parts can be cleaned with a paper towel and mild detergent It s not recommended to use rubbing alcohol for c...

Page 7: ...eplacement Fuse 3A 250V Cat No 1400 FS5 If the microscope is plugged in but the illuminator is not turning on the fuse could be blown The instructions below detail how to check for a blown fuse You ca...

Page 8: ...resistance do not force 4 Close the compartment door and follow the instructions in the Reflected Illumination Setup section to properly center the lamp Replacing the Reflected Illuminator Lamp Replac...

Page 9: ...lyzer knob located on the right side of the reflected illuminator must be pushed inward to engage this function figure 55 1 Carefully remove all components from foam packaging and plastic bags Use the...

Page 10: ...ve In addition you will need to fully open the aperture iris in order to prevent vignetting and leave the condenser height the same as when originally setup using the 10X objective 5 To control contra...

Page 11: ...er installed but should you need to reinstall at any point follow the directions below 13 Rotate the nosepiece so that either the 5X objective or an empty slot is in the light path 14 Using the coarse...

Page 12: ...making adjustments to the illuminator housing or lamp fixture 3 Ensure that the illumination selector knob is pushed all the way inward R figure 17 Figure 15 Figure 17 Figure 18 Figure 19 4 Rotate the...

Page 13: ...b adjusts for latitudinal X axis and the lower for longitudinal Y axis movement figure 21 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 10 Rotate each of these in succession until the field of vie...

Page 14: ...and should be checked and readjusted before each use Focus Adjustment 18 Focusing on your specimen is achieved through rotating the coarse and or fine focusing knobs These are located on either side o...

Reviews: