Transmitted Illumination Setup

23.

Set the main power switch to the ON position (“|”) and the illuminator selection
switch to the DOWN position (“|”) (figure 35). Adjust the VLC (variable lighting
control), located on the right side of the microscope base, counter-clockwise
to the maximum position.

24.

Check to see that the illuminator selector knob is pulled out all the way (“T”)
(figure 36).

25.

Rotate the field diaphragm dial clock-wise to fully open the iris (figure 37).

26.

Shift the aperture diaphragm slider to the left to fully open the iris (figure 37).

27.

Remove the transmitted polarizer filter from the collector if applicable. Pull
the  Analyzer selector knob fully out (figure 38).

28.

Rotate the 5X objective into the light path. You may have to lower the stage
in order not to crash the higher magnification objectives into the stage.

Continued on next page...

9.

Adjust the three reflected illuminator alignment knobs located on the
illuminator housing in order to properly align the lamp. The knob on the left
(figure 20) focuses the lamp transversely (Z-axis) while the upper right knob
adjusts for latitudinal (X-axis) and the lower for longitudinal (Y-axis)
movement (figure 21).

Figure 20

Figure 21

Figure 22

Figure 23

Figure 24

10.

Rotate each of these in succession until the field of view is at it’s brightest
overall and the brightest portion is centered in the field of view (figure 22).
This step may be aided by placing a bright white piece of paper on the
stage directly under the objective and the light field viewed on that.

Reflected Illumination Setup

 

(continued)

11.

Next, the reflected illuminator field diaphragm needs to be properly centered
in the light path. Ensure that the reflected illuminator aperture diaphragm is
fully open by rotating the Aperture dial to the right (figure 23).

Reflected Field Diaphragm Alignment

Continued on next page...

12.

While looking through the eyepieces close the field diaphragm by rotating
the Field dial to the left until you begin to see a darker perimeter in the field
of view resembling an iris silhouette(figure 24).

Figure 35

selector

power

focus

control

latitude

control

longitude

control

selector

knob

field iris

aperture

selector knob

29.

Look through the eyepieces and you should see a white field of view. You
may have to adjust the variable lighting control in case the light is too dim or
too bright. If the brightest portion of the field of view isn’t centered
(figure 39) you will need to perform the following instructions.

Summary of Contents for 1442MMi

Page 1: ...Printed in the USA VGMNL062512 1400MMi Operation Manual Model 1442MMi www veegee com 800 423 8842 sales veegee com support veegee com...

Page 2: ...croscope system Modifying the product in any manner for use not originally intended shall automatically void the manufacturer s warranty In no event shall VEE GEE Scientific Inc be held liable for any...

Page 3: ...surfaces are made Shock Hazard failure to follow instructions could lead to electrical shock Safety Precautions Before replacing the lamp or fuses always turn off the power switch and disconnect the p...

Page 4: ...ymptom Image through eyepieces is too dim Step Possible Cause HowToTest TestResult Solution For More Info 1 2 Variable lighting control VLC is turned down too far Visually inspect VLC is turned up VLC...

Page 5: ...turned down Go to next step Turn VLC up Page16 Microscope not getting power Inspect outlet Good outlet Outlet not good Go to next step Plug power cord into working outlet Blown fuse Visually inspect...

Page 6: ...an to prevent contaminate ingress is essential to the longevity of the objectives All other parts can be cleaned with a paper towel and mild detergent It s not recommended to use rubbing alcohol for c...

Page 7: ...eplacement Fuse 3A 250V Cat No 1400 FS5 If the microscope is plugged in but the illuminator is not turning on the fuse could be blown The instructions below detail how to check for a blown fuse You ca...

Page 8: ...resistance do not force 4 Close the compartment door and follow the instructions in the Reflected Illumination Setup section to properly center the lamp Replacing the Reflected Illuminator Lamp Replac...

Page 9: ...lyzer knob located on the right side of the reflected illuminator must be pushed inward to engage this function figure 55 1 Carefully remove all components from foam packaging and plastic bags Use the...

Page 10: ...ve In addition you will need to fully open the aperture iris in order to prevent vignetting and leave the condenser height the same as when originally setup using the 10X objective 5 To control contra...

Page 11: ...er installed but should you need to reinstall at any point follow the directions below 13 Rotate the nosepiece so that either the 5X objective or an empty slot is in the light path 14 Using the coarse...

Page 12: ...making adjustments to the illuminator housing or lamp fixture 3 Ensure that the illumination selector knob is pushed all the way inward R figure 17 Figure 15 Figure 17 Figure 18 Figure 19 4 Rotate the...

Page 13: ...b adjusts for latitudinal X axis and the lower for longitudinal Y axis movement figure 21 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 10 Rotate each of these in succession until the field of vie...

Page 14: ...and should be checked and readjusted before each use Focus Adjustment 18 Focusing on your specimen is achieved through rotating the coarse and or fine focusing knobs These are located on either side o...

Reviews: