autopilot
vmc-a601
140 N. Brandon
P.O. Box 74, Fallbrook, CA 92028
1-800-VANIMAN (826-4626)
VMC-a601.-I-09/12
Page 1
Air Handpiece Users
Connect the air turbines by connecting the air hoses and tees provided on the
AutoPilot
to the air hos-
es that go to the air turbine. Cut the air turbine air hose betwen the foot pedal and turbine and recon-
nect the hose using the tee provided on the
AutoPilot
. When either of the two air turbines is started,
the dust collector should start automatically. When the last air turbine (or micromotor) stops, the dust
collector will stop after a 5 second delay.
NSK Z500 Series Handpiece Users
From the NSK Z500 Users manual:
“Vacuum-coupled Mode
On some dental tables with vacuum dust collector, the motor may be used while being coupled with a
dust collector.
When such a dust collector is used, power consumption of the Z500 can be regulated so that the
vacuum coupled function can work. If you need coupling with a vacuum dust collector, select the
mode as follows:
How to select the mode
While pressing Forward/Reverse selector switch (button), turn on the Power switch, and the mode can
be selected. A long beep indicates vacuum-coupled mode and 2 short beeps indicate non-coupled
mode. each time the switch selection is made, the mode changes between vacuum-coupled mode
and non-coupled mode.”
troubleshooting
Problem: Unit does not turn on dust collector when micromotor is turned on.
Causes:
1. Check that micromotor is plugged in (see instructions above).
2. Test dust collector with standard wall socket to see that it is operating correctly.
3. Check circuit breaker and push to reset.
Problem: Micromotor does not turn on.
Causes:
1. Check circuit breaker and push to reset.
2. Check that the
AutoPilot
plugged in correctly.