background image

autopilot
vmc-a601

140 N. Brandon

  

  

P.O. Box 74, Fallbrook, CA 92028

  

  

1-800-VANIMAN (826-4626)

VMC-a601.-I-09/12

Page 1

Air Handpiece Users

Connect the air turbines by connecting the air hoses and tees provided on the 

AutoPilot 

to the air hos-

es that go to the air turbine.  Cut the air turbine air hose betwen the foot pedal and turbine and recon-

nect the hose using the tee provided on the 

AutoPilot 

.  When either of the two air turbines is started, 

the dust collector should start automatically.  When the last air turbine (or micromotor) stops, the dust 

collector will stop after a 5 second delay.

NSK Z500 Series Handpiece Users
From the NSK Z500 Users manual:

“Vacuum-coupled Mode
On some dental tables with vacuum dust collector, the motor may be used while being coupled with a 

dust collector.

When such a dust collector is used, power consumption of the Z500 can be regulated so that the 

vacuum coupled function can work. If you need coupling with a vacuum dust collector, select the 

mode as follows:

How to select the mode

While pressing Forward/Reverse selector switch (button), turn on the Power switch, and the mode can 

be selected. A long beep indicates vacuum-coupled mode and 2 short beeps indicate non-coupled 

mode. each time the switch selection is made, the mode changes between vacuum-coupled mode 

and non-coupled mode.”

troubleshooting

Problem:  Unit does not turn on dust collector when micromotor is turned on.
Causes:  
 

1. Check that micromotor is plugged in (see instructions above).

 

2. Test dust collector with standard wall socket to see that it is operating correctly.

 

3.  Check circuit breaker and push to reset.

Problem:  Micromotor does not turn on.
Causes:
 

1. Check circuit breaker and push to reset.

 

2. Check that the 