Vapor Expert(PY-001)Portable Nebulizer – Instruction Manual 

 

 

 

When the power button is pressed and the appliance starts working the green light 
will on (sometimes, in orange and which depend on battery power remained) and the 
build in timer will start to countdown. 

 

If the indicator turns orange, it shows the battery is weak.

 

If the appliance is activated 

and the power indicator turns to red, it means the batteries are exhausted and the 
appliance cannot operate properly. Please replace with new battery immediately.   

 

If there is no medicine inside the medical cup, indicator light will turn red and flash 
four times before appliance turning off automatically. 

 

CLEAN AND MAINTENANCE NOTICE 

After each using, please clean the following parts: 

1.

 

Remove the medical cup and mouthpiece or mask. Discard all 
remaining medicine in the cup. 

 

Open the medical cap and discard all remaining medicine. 

 

Pour in some hot water and close the cover completely. 

 

Press the power button and operate 1~2minutes to discard all 
remaining medicine completely. 

     

Ensure medicine be cleaned completely after each using, otherwise, 

it could cause mesh blocked. 

                               

                                                                     

 

2. Clean the surface of main body of appliance with cotton swabs or a piece 

of fine cotton fabrics. 

Please clean the electrode in the main body and medical cup, ensure                   
normally electrical conduction. 

Please do not use easily volatility liquid to clean the electrode, ex. 

Benzene, thinner. 

     

         

3. Clean the medical cup, mouthpiece and mask with clean water and store at a safety 

place after cleaning the surface with ethanol. 

Suggestion: Use a new medical cup for different illness or patients to avoid cross infection. 

Summary of Contents for PY-001

Page 1: ...Vapor Expert PY 001 Portable Nebulizer English 2011 10 21 Instruction Manual ...

Page 2: ...under the guidance of qualified medical experts Environment This product is intended for use in a medical facility such as Hospital clinic and doctor s office in a room of general household and in open air environment with a roof Durable period Durable periods are as follows Durable period may vary depending on usage environment Main body 1year Medical cup 1year Mouth piece 1year Mask 1year Precau...

Page 3: ...asily to carry 3 Ultrasonic vibration which produces 100 000 120 000 vibrations per second 4 After 20 minutes continuously operation or when the medical cup is no medicine the appliance will automatic shutdown 5 Built in low battery indicator 6 No matter use the product in different angle that still can operate normally in short term When put the product upside down medicine cannot contact to mesh...

Page 4: ...en the power is good for working When the power light turns to orange the battery is become weak but you still can use product however you may need to replace battery soon If the power light is turns red it means voltage is too low to operate please replace the battery immediately During using this appliance if any irregularity or discomfort is observed please discontinue use this appliance This p...

Page 5: ...lso showing is there still have medicine inside the medical cup If there is no medicine inside indicator light will turn red and flash four times before turn off automatically 3 Medical Cup Contain the medicine Suggestion Replace a new medical cup per year 4 Push button push here to release the medical cup from the main body 5 Medical cup cover open here to pour the medicine 6 Mouthpiece or Mask 7...

Page 6: ...ied by using different brands of battery Suggestion Use Alkaline batteries or Ni MH rechargeable Batteries option 1 Remove batteries if the unit is not in use for long periods Failure to do so could result in damage due to battery leakage 2 Discard the waste batteries according to local law 3 Open the medical cup cover and pour the medicine into the medical cup and close the cover completely 4 Ins...

Page 7: ...n the following parts 1 Remove the medical cup and mouthpiece or mask Discard all remaining medicine in the cup Open the medical cap and discard all remaining medicine Pour in some hot water and close the cover completely Press the power button and operate 1 2minutes to discard all remaining medicine completely Ensure medicine be cleaned completely after each using otherwise it could cause mesh bl...

Page 8: ... red indicator light is on the battery power is too low Please replace the battery immediately 2 Discard the waste batteries according to local law 1 Check the battery is installed correctly Open battery cover and check if the battery electrode is rusted or not Cannot nebulize or not enough mist output 1 Check if the medical cup is filled with appropriate amount of medicine 2 The mesh in medical c...

Page 9: ...anual The accessories Model No Size Child Mask PY 001 MS 87 22 x 64 35 x 57 mm Adult Mask PY 001 ML 124 x 77 x 68 5 mm Mouth Piece PY 001 MP 43 x 27 5 mm Medical Cup Gray PY 001 MC 54 x 43 99 x 38 50 mm Medical Cup Blue PY 001 B MC 54 x 43 99 x 38 50 mm ...

Page 10: ...Vapor Expert PY 001 Portable Nebulizer Instruction Manual ...

Page 11: ...Vapor Expert PY 001 Portable Nebulizer Instruction Manual ...

Page 12: ...therapy equipment Part 1 Nebulizing systems and their components Annex CC 3 using the type test method Particle Size MMAD 4 98um Sound Less than 30db Aerosol Output 0 2 mL min SYMBOLS DESCRIPTION Warning Symbol for Type BF applied part Direct current Read the documentation symbol Batch Symbol Serial number Symbol Manufacturer EC Representative Symbol ...

Page 13: ...ght Low battery indication 2 0V red light Operation environment 10 40 C 30 80 RH Storage Transportation environment 20 70 C 85 RH Nebulization rate 0 2mL min 1 0mL min Max Particle size 5um Time AUTO OFF Power on operate 20 min Then auto off Nebulizer auto off No medication Medication bottle volume 2 8 mL Max Ultrasonic Oscillation Frequency 103 KHZ 123 KHZ 1K Reusable parts life Medical cup mask ...

Page 14: ...ch Co Ltd 2F No 285 Sec 2 Tiding Blvd Neihu Dist Taipei City 11493 Taiwan R O C Tel 886 2 2627 0630 Fax 886 2 2627 0636 E mail service puyuan bio com Ms Chien Wei Lin Breitachstr 56 D 73547 Lorch Germany Tel 49 7172 9350580 Fax 49 7172 9350589 E Mail linchienwei gmail com ...