A-4
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
Immunity
test
0% Uτ
(100% dip in
Uτ) for 0.5
cycle
0% Uτ
(100% dip in
Uτ) for 1 cycle
70% Uτ
(30% dip in Uτ)
for 25/30
cycles
0% Uτ
(100% dip in
Uτ) for 5 s
0% Uτ
(100% dip in
Uτ) for 0.5
cycle
0% Uτ
(100% dip in
Uτ) for 1 cycle
70% Uτ
(30% dip in Uτ)
for 25/30
cycles
0% Uτ
(100% dip in
Uτ) for 5 s
IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
Mains power quality
should be that of a
typical commercial
or hospital
environment. If the
user of the
MEDIC
Vaporizer requires
continued operation
during power mains
interruptions, it is
recommended that
the
MEDIC Vaporizer be
powered from an
uninterruptible
power supply.
Power
Frequency
Magnetic Field
(50/60 Hz)
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Note
: Uτ is the a.c. mains voltage prior to application of the test level.
INFORMATION ON ELECTROMAGNETIC COMPATIBILITY