Summary of Contents for 2000i

Page 1: ...2000i Operating Instruction Manual...

Page 2: ...his manual Vapotherm also recommends that users follow the Centers for Disease Control CDC publications Guidelines for Maintenance of In Use Respiratory Therapy Equipment and Guidelines for Prevention...

Page 3: ...be used to deliver pressure in a closed system 2 3 Cautions Verify that the power source is compatible with the electrical specifications shown on each component For proper grounding reliability conne...

Page 4: ...w flow cartridge VT01 BS Ensure that the correct cartridge is inserted before operating Section 3 About the Vapotherm 2000i The Vapotherm 2000i warms and humidifies flows of air oxygen or medical gas...

Page 5: ...fer Cartridge 1 The Vapor Transfer Cartridge VT01 AS or VT01 BS attaches to the unit by two water and two air connections 2 When facing the unit access is via a hinged cover on right side The cartridg...

Page 6: ...rtridge or Patient Delivery Tube and both should be checked or changed NOTE If using a low flow cartridge VT01 BS the flow cannot be decreased below 5 lpm until an appropriate cannula has been attache...

Page 7: ...Check that water is properly circulating through the machine by making sure the Patient Delivery Tube is warm across the entire length 2 If good circulation cannot be confirmed check that the water f...

Page 8: ...Manual Page 14 If further assistance is needed please call your clinical product specialist or local distributor representative S S If the Cartridge Alarm is continuous and air bubbles are rising int...

Page 9: ...approved reprocessing area outside the patient care area Warning The water circuit of the Vapotherm 2000i System is not sterile and can potentially have bacterial contamination The water circuit of th...

Page 10: ...70 90 isopropyl alcohol disinfectant wipes when wiping down the Vapotherm 2000i and disinfection kit components Always use a new disinfectant wipe taken directly from the package or container and ens...

Page 11: ...ocol 8 5 Disinfect Gas and Water Circuits 1 Open clamp on Bag A Disinfectant will start to drain from Bag A through the unit and into Bag B Fig 9 2 When disinfectant has stopped draining into Bag B cl...

Page 12: ...ral Information 9 1 Specifications Dimensions Height 11 280 mm width 5 5 140 mm depth 4 5 114 mm excluding IV pole clamp Weight Less than 6 lbs 2 7 kg without water reservoir Vapotherm spike set Works...

Page 13: ...Appendix A WARNINGS These disinfectant solutions are designed to be used to disinfect the Vapotherm 2000i machine without the cartridge in place These disinfectant solutions ARE NOT approved to disinf...

Page 14: ...syringe or pipette graduated in 1ml increments Concentrate Disinfectant Minncare Cold Sterilant EPA Reg No 52252 4 22 00 Hydrogen Peroxide 4 50 Peroxyacetic Acid 73 50 Inert Ingredients Control III Di...

Page 15: ...Vapotherm Inc 198 Log Canoe Circle Stevensville MD 21666 T 001 410 604 3977 F 001 410 604 3978 www vtherm com PN 8 300068 00 Rev E bdc 8369 Supplied by...