background image

Operating Instruction Manual 

Page 13

Alarm 

Cause 

Action 

indication

NOTE: To restart after a system failure, the unit must be reset by a momentary pressure on the Power 
button. Do not hold the Power button. The alarm will shut off after a delay of about a second, and 
the unit can then be restarted normally.

S

S

Water is not filling system properly

 

 

Low Water Pressure

 

 

Malfunctioning Water or Gas

 

Pressure Sensor

Make sure VSS-1 spike set is open, and

 

the tube is not kinked or blocked by 

air bubbles.

Make sure gas and water connections are 
open, gas can flow to unit, and air has 
purged from water system: if not, run 

‘CLEANING MODE’.

Send in for service.

Water low

Insufficient gas or water pressure.

 

 

 
 

Malfunctioning 

Water or Gas

 

Pressure Sensor

Make sure the gas and water circuits are 
open and functional and air has purged 
from the water circuit if the unit is in 

normal operating mode.

Run in “CLEANING MODE”.

Make sure there is correct flow for 

cartridge flow rates. If using <5lpm with a 

low flow cartridge, a nasal cannula must 

be attached.

If a component failure return the unit 

for service.

System Failure

Cartridge

Water drops in the circuit will 
cause a cartridge alarm; this does 

not necessarily mean the cartridge 

needs to be replaced. 

Section 6

 Alarms, Trouble Shooting and Component Change-Outs

6.1

  General

1.   Periodically check for alarm conditions.

2.   Unit will shut down if there is no gas flow. However, flow will not be interrupted if unit  

shuts down or malfunctions for any other reason.

3  Unit will shut down if temperature safety limits are exceeded, or if water level is low for  

more than 4 mins. However, unheated gas flow will continue.

NOTE: 

Should a malfunction occur, indicators on the front panel will light and an alarm will sound. 

If the actions listed here do not correct the problem causing the alarm, the unit should be returned to 
an approved facility for service.

6.2

  Alarms and Troubleshooting

Section 5 

Operation of The Vapotherm

 2000i

5.7

  Operations – General Guidelines

1.  Check that water is properly circulating through the machine by making sure the Patient Delivery 

Tube is warm across the entire length.

2.  If good circulation cannot be confirmed, check that the water flow is not obstructed by air bubbles.
3.  Take precautions to minimize cooling of the unheated cannula by trying to maintain contact with 

the patient’s skin and insulating the exposed portion of the cannula with bedding.

4.  Cartridge door should be closed during operation.

NOTE: Condensation in the cannula may occur at low flow rates. To minimize condensation, these 
general guidelines should be followed:

• 

If using flow rates less than 5 lpm, do not set the temperature higher than 34ºC.

•  The Vapotherm unit should not be in a position where it is cooled (eg. by an air conditioning outlet). 

CAUTION:

 DO NOT EXCEED flows of 8 lpm for VT01-BS and 40 lpm for VT01-AS cartridge. 

DO NOT SET flows below 1 lpm for VT01-BS.

WARNING:

Never occlude the nares with cannula.

NOTES:

It may become necessary to disconnect the cannula from the Patient Delivery Tube for short periods, such as 
when moving a patient out of a radiant warmer. At flow rates less than 5 lpm, cannula disconnection will 
activate a system failure alarm, requiring a reset. To avoid this alarm, briefly turn off unit by pressing the power 
key once. The display will show two bars. Disconnect the cannula from Patient Delivery Tube and move the 
patient, reconnect the cannula, then press the power key once more to restart the unit.
A SYSTEM FAIL (88) alarm will activate if there is insufficient gas pressure in the manifold. If no cannula is fitted, 
flow rate at startup should be at least 5 lpm. The minimum flow rate for operation is 1 lpm if a neonate, infant 
or premature-sized cannula is fitted, 5 lpm with a pediatric cannula, and 8 lpm with an adult cannula.
Cannulas are single use patient items, dispose of as necessary or according to your institution’s guidelines or  
when visibily soiled or excessively wet from secretions. Weekly change out is recommended to avoid any 
hardening of nasal prongs.
An air lock can develop at the pump, preventing normal water flow. Try restarting the unit in 
cleaning mode.

Operating Instruction Manual 

Page 12

Summary of Contents for 2000i

Page 1: ...2000i Operating Instruction Manual...

Page 2: ...his manual Vapotherm also recommends that users follow the Centers for Disease Control CDC publications Guidelines for Maintenance of In Use Respiratory Therapy Equipment and Guidelines for Prevention...

Page 3: ...be used to deliver pressure in a closed system 2 3 Cautions Verify that the power source is compatible with the electrical specifications shown on each component For proper grounding reliability conne...

Page 4: ...w flow cartridge VT01 BS Ensure that the correct cartridge is inserted before operating Section 3 About the Vapotherm 2000i The Vapotherm 2000i warms and humidifies flows of air oxygen or medical gas...

Page 5: ...fer Cartridge 1 The Vapor Transfer Cartridge VT01 AS or VT01 BS attaches to the unit by two water and two air connections 2 When facing the unit access is via a hinged cover on right side The cartridg...

Page 6: ...rtridge or Patient Delivery Tube and both should be checked or changed NOTE If using a low flow cartridge VT01 BS the flow cannot be decreased below 5 lpm until an appropriate cannula has been attache...

Page 7: ...Check that water is properly circulating through the machine by making sure the Patient Delivery Tube is warm across the entire length 2 If good circulation cannot be confirmed check that the water f...

Page 8: ...Manual Page 14 If further assistance is needed please call your clinical product specialist or local distributor representative S S If the Cartridge Alarm is continuous and air bubbles are rising int...

Page 9: ...approved reprocessing area outside the patient care area Warning The water circuit of the Vapotherm 2000i System is not sterile and can potentially have bacterial contamination The water circuit of th...

Page 10: ...70 90 isopropyl alcohol disinfectant wipes when wiping down the Vapotherm 2000i and disinfection kit components Always use a new disinfectant wipe taken directly from the package or container and ens...

Page 11: ...ocol 8 5 Disinfect Gas and Water Circuits 1 Open clamp on Bag A Disinfectant will start to drain from Bag A through the unit and into Bag B Fig 9 2 When disinfectant has stopped draining into Bag B cl...

Page 12: ...ral Information 9 1 Specifications Dimensions Height 11 280 mm width 5 5 140 mm depth 4 5 114 mm excluding IV pole clamp Weight Less than 6 lbs 2 7 kg without water reservoir Vapotherm spike set Works...

Page 13: ...Appendix A WARNINGS These disinfectant solutions are designed to be used to disinfect the Vapotherm 2000i machine without the cartridge in place These disinfectant solutions ARE NOT approved to disinf...

Page 14: ...syringe or pipette graduated in 1ml increments Concentrate Disinfectant Minncare Cold Sterilant EPA Reg No 52252 4 22 00 Hydrogen Peroxide 4 50 Peroxyacetic Acid 73 50 Inert Ingredients Control III Di...

Page 15: ...Vapotherm Inc 198 Log Canoe Circle Stevensville MD 21666 T 001 410 604 3977 F 001 410 604 3978 www vtherm com PN 8 300068 00 Rev E bdc 8369 Supplied by...

Reviews: