Page 8
FiO2
(%)
Maximum Precision Flow Rate (L/min)
Sea
Level
1,000
ft
2,000
ft
3,000
ft
4,000
ft
5,000
ft
6,000
ft
7,000
ft
8,000
ft
21%
40
40
40
39
37
34
32
30
28
22%
40
40
40
39
37
35
33
30
28
23%
40
40
40
40
38
35
33
31
29
24%
40
40
40
40
38
36
34
31
29
25%
40
40
40
40
38
36
34
31
30
26%
40
40
40
40
39
36
34
32
30
27%
40
40
40
40
39
37
35
32
30
28%
40
40
40
40
40
37
35
33
31
29%
40
40
40
40
40
38
36
33
31
30%
40
40
40
40
40
38
36
33
31
31%
40
40
40
40
40
39
37
34
32
32%
40
40
40
40
40
39
37
34
32
33%
40
40
40
40
40
40
38
35
33
34%
40
40
40
40
40
40
38
35
33
35%
40
40
40
40
40
40
38
36
34
36%
40
40
40
40
40
40
39
36
34
37%
40
40
40
40
40
40
40
37
34
38%
40
40
40
40
40
40
40
37
35
39%
40
40
40
40
40
40
40
38
35
40%
40
40
40
40
40
40
40
38
36
41%
40
40
40
40
40
40
40
39
36
42%
40
40
40
40
40
40
40
39
37
43%
40
40
40
40
40
40
40
40
38
44%
40
40
40
40
40
40
40
40
38
45%
40
40
40
40
40
40
40
40
39
46%
40
40
40
40
40
40
40
40
39
47%
40
40
40
40
40
40
40
40
40
48%
40
40
40
40
40
40
40
40
40
49%
40
40
40
40
40
40
40
40
40
50% or
Greater
40
40
40
40
40
40
40
40
40
3100772 Rev E
Summary of Contents for Q50
Page 1: ...Q50 Compressor Instructions for Use...
Page 2: ......