Power Inlet Switch
The Vapotherm Q50 compressor makes use of a medical grade
power inlet switch. The power inlet switch contains two (2) fuses.
• 5ET 8-R (5mm X 20mm, 8A, SLOW BLOW, GLASS FUSE,
250V)
WARNING: To avoid risk of electric shock, this equipment
must only be connected to supply mains with
protective earth.
Air Intake
The compressor utilizes air from its surrounding environment. It
passes through multiple stages of filtration prior to delivery to the
Precision Flow. The first filtration stage is accessible from the outside
and should be cleaned once per week according to maintenance
procedures. Further filtration is accessible only by trained personnel
during preventative maintenance.
WARNING: Never restrict or block the air intake. This could
result in failure to deliver therapy as intended.