Operating Instructions
Pre-operational Check
Prior to starting the Q50 unit, the user should verify the following:
1. Check the power supply
Ensure that the power cord is installed into an outlet with the
appropriate power requirements.
2. Check that the Precision Flow unit is powered off
Disconnect the Precision Flow power.
3. Check the outlet hose to the Precision Flow unit
Verify that the outlet hose that supplies the Precision Flow unit
with air is fully connected to the compressor and Precision
Flow unit before turning the unit on.
4. Connect oxygen supply to Precision Flow (if applicable)
Connect an oxygen supply as directed in the Precision Flow
Instructions for Use if oxygen concentrations above 21% are
desired.
WARNING: Verify that the water bottle reservoir and water
trap have been emptied prior to starting the compressor.
Failure to do so may result in the spillage of liquid and
could pose a slip hazard.
WARNING: Verify that the air intake on the back of the
unit
and air exhaust on the bottom of the units are
unobstructed.
NOTE: When connecting to a wall air source, use the manual
draining water trap and gas hose provided with the
Precision Flow.