Cleaning and Disinfection
This manual is intended only to give general guidelines for the
cleaning and disinfection of the compressor. It is the end user’s full
responsibility to ensure that any methods and techniques used are
effective and conform to the institution’s guidelines and procedures.
Part
Procedure
Comments
Compressor
Exterior
Wipe the exterior
according to hospital
disinfection policies.
Use clean water to
remove any residual
residue.
Do not allow liquids
to infiltrate beyond
the exterior surfaces
of the compressor,
or into any electrical
interfaces or
components.
Water
Bottle
Reservoir
Wipe the exterior of the
water bottle reservoir
and reservoir tubing
according to hospital
disinfection policies
after emptying the
water bottle reservoir
and in between patient
uses.
Do not allow liquids
to infiltrate beyond
the exterior surfaces
of the compressor,
or into the interior
foam housing of the
compressor.
Rinse the water bottle
reservoir with warm
soapy water after
emptying the water
bottle reservoir and in
between patient uses.
Confirm rinsing of
the water bottle
reservoir to remove
soap residue.
Page 22
Air Intake Filter
The air intake filter is located on the back of
the unit. It should be removed once a week
and cleaned. Brush any gross dust particles
off the surface of the filter. Submerge the filter
in warm soapy water and gently squeeze and
massage the filter to flush any dust particles
out of the interior of the filter. Air dry.