Maintenance
The following maintenance should be completed at the stated
intervals. Failure to do so could result in underperformance of the
system and/or a reduced life of the product.
• Air intake filter cleaning once per week (wash). Replace every
4,000 hours or if there are any signs of tears, degradation or
clogging of the filter material.
• Replacement filter for the auto draining air filter once per
4,000 hours
• Air intake muffler replacement once per 4,000 hours
• Compressor rebuild once per 8,000 hours
• Compressor motor replacement once per 20,000 hours
• Replace the filter element (Vapotherm Part Number 3101397)
within the auto-draining water trap every 6 months or if any
visible signs of contamination are present.
For ordering information please contact Vapotherm Customer
Service at 1-866-566-2652.
WARNING: Disconnect power supply prior to opening unit.
Failure to do so may result in serious injury.
CAUTION: Unit should only be serviced by qualified and
trained personnel.