Trouble
Fault State
Possible Cause Method
No air
output,
cont.
The cooling
fan is running
but not the
compressor
Compressor has
overheated
Ensure that
the air intake
filter is clean
and the air
intake and air
exhaust are not
blocked. Allow
the compressor
to cool for
20-30 minutes
and attempt to
restart
The cooling
fan is running
but not the
compressor
even when the
compressor is
cool
The compressor
has failed
Call for service
The compressor
and cooling fan
are running
The air pipeline
is not tightened
properly
Tighten the
fittings
The air pipeline
hose is OK and
the exhaust
noise can be
heard inside the
machine
Pipeline
inside has
disconnected.
Call for service
Air output is
low causing
PF to alarm
Insufficient
pressure
Internal pressure
regulator below
recommended
operating levels
Call for service