Quick Start Guide
Review all instructions and warnings prior to use.
1. Ensure the Precision Flow unit is unplugged and the power outlet
meets the appropriate power requirements.
2. Mount the compressor as needed. See “Mounting Positions,”
page 13.
3. Install and connect the High Altitude / High Flow Kit. Remove the
water trap from the air port on the back of the Precision Flow
and Install the auto-draining water trap assembly on the Precision
Flow.
4. Attach the end of the air hose onto the DISS air fitting on the back
of the Q50 Compressor and hand tighten.
5. Attach the other end of the air hose to the DISS air fitting on the
air filter assembly on the back of the Precision Flow unit and hand
tighten.
6. Connect an oxygen supply to the DISS oxygen fitting on the
oxygen filter assembly on the back of the Precision Flow, if
oxygen concentrations above 21% are desired, and pressurize
the oxygen source.
7. Attach the Q50 power cord to the compressor power cord inlet
and plug into the power outlet.
8. Toggle the power inlet switch into the “I” position. The Q50
Compressor will now run.
9. Plug the Precision Flow into the power outlet and press the
run/standby button to begin start up.
WARNING: Verify that the water bottle reservoir has been
emptied prior to starting the compressor. Failure to do so
may result in the spillage of liquid and could pose a slip
hazard.
WARNING: Do not block or restrict the air exhaust
underneath the compressor. This could result in failure to
deliver therapy as intended. When the unit is placed on the
floor, be sure that any flooring, such as carpet, does not
restrict the air exhaust.