SET-UP INSTRUCTIONS

Ensure the Precision Flow and Q50 are unplugged.
Mount the compressor as needed.
Install and connect the High Altitude / High Flow Kit:
  Remove the water trap from the air port on the back of the Precision Flow
  Install the auto-draining water trap assembly on the Precision Flow.
  Connect the free end of the air hose to the DISS fitting on the Q50  

 Compressor.
Connect oxygen supply, if needed, and pressurize.
Plug in and toggle the on/off switch on the back of the Q50 compressor.
Plug in the Precision Flow.

PRE-OPERATION CHECK

Check or verify that:

The power supply is appropriate for Q50 power requirements.
The Precision Flow is powered off and unplugged.
The air hose is securely connected to the Q50 and the Precision Flow.
The  oxygen  hose,  if  needed,  is  connected  to  an  oxygen  source  and  the 

Precision Flow, and the source is pressurized.
The Q50 water bottle and the Precision Flow water traps are empty.
The Q50 air intake on the back of the unit and the exhaust on the bottom of 

the unit are unobstructed.

RUNNING

Check the water bottle and Precision Flow water traps once per shift.
Remove and clean air intake filter once per week.