2920 Float & Tape Transmitter
5 Display and Configuration Interface
80
Installation and Operations Manual
15. Enter values 0 to 100 (percent) and verify the measurement on the 4-20 mA Meter:
0.0 = 4.00 +/- 0.01
25.0 = 8.00 +/- 0.01
50.0 = 12.00 +/- 0.01
75.0 = 16.00 +/- 0.01
100.0 = 20.00 +/- 0.01
BPM Data
Note
The BPM (Bi-Phase Mark) Data menu is used only when monitoring communica-
tions or when configuring a Bi-Phase Mark Device.
The BPM Data menu contains parameters and allows configuration of the BPM data point.
BPM Data menu presents 3 options:
• Display Only — data display, such as message and number of requests
• Config Params — parameters that are used to configure data in BPM data, such as
response delay
• All Params — scrolls through the Display, Commands, and Config parameters in one
menu
To Access BPM Data:
1. Press Enter to access the Main Menu. The System Menu point is automatically highlighted.
2. Press Minus ten times to move the highlight to BPM Data.
3. Press Enter and the BPM Data menu opens.
4. To select one of the options, press Minus to move the highlight to the option you want and
then press Enter.
Refer to the table below for more information on each BPM Data option.
Display Only
Commands
Config Params
All Params
Main Menu
BPM Data
NMT Prothermo
HART Device
10
BPM Data
Summary of Contents for 2920
Page 2: ......
Page 16: ...2920 Float Tape Transmitter 1 Introduction 6 Installation and Operations Manual...
Page 114: ...2920 Float Tape Transmitter 6 Bi Phase Mark 104 Installation and Operations Manual...
Page 120: ...2920 Float Tape Transmitter 7 MODBUS 110 Installation and Operations Manual...
Page 126: ...2920 Float Tape Transmitter 9 L J TankWay 116 Installation and Operations Manual...
Page 158: ...2920 Float Tape Transmitter 14 Ordering Information 148 Installation and Operations Manual...
Page 193: ......