Liquid Level Indicator
22
Installation and Operations Manual
Initial Calibration
These instructions apply to systems that use a hollow-shell, welded float that is 14.5 inches
[368 mm] in diameter by 4.0 inches [102 mm] deep. Calibration is based upon the standard
specific gravity of water (1.0).
Note
For products of other specific gravities to be gauged, adjust the indicator accord
-
ingly. Refer to the industry related charts for the particular product.
With the float just touching the bottom of the tank, water will be at the float equator (water
line). Use the float connecting clamps (see Figure 7 on page 16) to make the final adjustment,
then trim excess tape or cable at the float end, leaving some additional tape or cable for future
adjustment needs.