USER
M
ANUAL
www.vareximaging.com
29
2315NDT v2 CMOS X-ray Detector User Manual
To make sure there is no overlap of the T
XRAY
with the frame readout, it is recommended to have
T
XRAY
configured shorter than T
EW
, so as to allow some delay time prior to X-ray
exposure and some tail time afterwards.
In some generators, there may be a delay between asserting Sync Out 2 and the actual start of
X-ray emissions. This must be compensated by increasing the programmed exposure window
T
EW
, so that the frame readout does not start before the X-ray emission has finished.
The frame readout time and fps of 2315NDT detectors are shown in
shows an example of the Software Trigger configuration.
Figure 9
Sync Connection for Software Trigger
In this case, the generator is informed that the detector is ready to receive the X-ray emission
when the Sync Out level changes.
Table 13
fps and Frame Readout Time for 2315NDT
2315NDT Acquisition speed
Binning
Binning mode
Frame readout Time
(T
R/O
)
2315NDT
fps
1×1
-
37.77 ms
26
2×2
Analog
14.25 ms
70
2×2
Digital
33.22 ms
30
4×4
Analog
11.69 ms
85
4×4
Digital
33.22 ms
30