User Manual: UM50566-EN
Optica 30/40/50 series
Revision: 3.0
Date of release: 2019-01-09
12/36
www.vareximaging.com
2.5.
Known contraindication(s)
Contraindications of the X-ray system continue to apply (see X-ray system documentation).
Additional the Optica 30/40/50 series introduces matter in the primary X-ray beam which primarily
results in additional X-ray scatter. Additional X-ray scatter causes a radiation risk. The Optica 30/40/50
series is mitigating the radiation risk by shielding the additional scatter and limiting the size of the
primary X-ray beam and also confining the scatter generated from the X-ray tube (housing).
2.6.
Restrictions on use
RF communication devices should be kept at a minimum distance of 30 cm. Otherwise the
performance of the device can be negatively influenced.
2.7.
Overview of the collimator
A
Fixing flange
B
Control knob, Cross shutters (Optica 30/40 Series only)
C
Control knob, Long shutters (Optica 30/40 Series only)
D
Enclosure
E
Status LED (Optica 30/40 Series only)
F
Push button for activation of the Light field and the Bucky laser (Optica 30/40 Series only)
G
Measuring tape, for determining the SID.
Pull out the end hook and place it against the designated surface (e.g. table, wall stand, etc.).
Read the SID size at the measuring tape aperture of the enclosure front.
H
Bucky laser shutter to close the laser aperture, see §2.8
I
Accessory rails, for a maximum of 2 accessories with a width of 177 mm
J
Accessory lock, for locking the accessory
Prior to inserting accessories in the accessory rails check the fit and locking.
Hazard of falling accessory.
K
Touch screen (Optica 40 Series only)