1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
10
www.vareximaging.com
U
SER
M
ANUAL
8.3.2
Connecting the Cables
To connect the cables, perform the following steps.
provides a graphical representation
of how a typical system setup appears.
1
Connect the
Mains
to the
Power Supply.
2
Connect the
Power Supply
to the
Detector Cable
.
3
Connect
Receptor Sync to Generator
to the
generator interface
.
4
Connect the
Detector Cable
to the
Power connector
on the detector.
5
Connect the
Ethernet cable
between a
Gigabit capable port
on the host computer and the
detector. The use of a Hub or switch is neither recommended or supported.
6
Connect the
Ground Cable
to the grounding lug (not shown in
7
Connect the
Ground Cable
to an acceptable ground connection.
8
Connect the
Mains Cable
to an available
power outlet
.
WARNING
Replacement power supply cords may use inadequately-rated wiring
and could damage the detector. If a replacement power supply cord is
required, please request one directly from Varex Imaging to ensure
that the power cord meets the standards for the detector.
Note
The X-ray generator interface is user-supplied equipment.