8.7
Environmental Considerations
Environments outside the specification reduce the lifetime and may irreparably damage the
Product.
8.8
Interface Signals
The synchronization interface to the detector consists of one opto-isolated input signal and one
opto-isolated output signal. Connection to these two signals can be made via the 8 pin LEMO
connector on the detector or via 9 pin D-Sub connector.
8.8.1
User_Sync Input Signal
The
User_Sync
input is used to trigger detector readout if the detector is in
User_Sync
mode
as shown in
The detector can be configured to do the following:
•
Respond to
User_Sync
pulse width, see
•
Not respond to
User_Sync
pulse width, see
The detector will respond with the actions in
when
User_Sync
Blanking is activated for
that particular mode.
Table 11
Environmental Considerations
Category
Limits
Storage & Transport (ambient)
-20º C to +70º C
Storage Humidity Range (non-condensing)
10% to 90%
Normal Operation Temperature (measured at the center of the
back cover)
10º C to 35º C
Operation Humidity (non-condensing)
10% to 90%
Atmospheric Pressure Range
700hPa to 1060hPa
Altitude
< 3000m
Table 12
User_Sync Input Activation
Input
Voltage
Current
V
in
minimum
4V
5mA
V
in
maximum
15V
25mA
Note:
V
in
minimum needs to be provided to the detector’s opto-coupler input.
Note:
V
f
is typically 1.2V.