1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
18
www.vareximaging.com
U
SER
M
ANUAL
User_Sync
Blanking is activated/de-activated through a register in the detector configuration
file. In both cases, the detector readout is triggered after the rising edge of the
User_Sync
pulse.
The minimum
User_Sync
pulse width is 200µs.
Figure 11
User_Sync Blanking Activated
Table 13
User Sync Pulse Duration and Response
User_Sync Pulse Duration
User_Sync Pulse Response
Pulse Duration < 200µs
User_Sync
signal is ignored
200µs < Pulse Duration < 4ms
Detector is readout and Valid Image bit is set to 1 in
Diagnostic Data. VCP ignores the image data.
Pulse Duration
>
4ms
Detector is readout and Valid Image bit is set to 0
and the frame is handled normally.
Note
The detector readout time may vary and depends on the detector mode.