1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
Interface and Power Supply and Cables
Caution
See
“Section 8.7, Environmental Considerations” on page 17
for
temperature and humidity levels. This may necessitate air flow over
the back surface of the detector.
Caution
An Image Quality Test or Customer Acceptance test should be
performed before the detector is used for the first time.
Note
The detector is designed to be integrated into a complete X-ray system
by a qualified system integrator.
WARNING
Be sure to turn OFF the power of the X-ray Detector, including turning
off the power supply before servicing, maintaining, connecting, or
disconnecting the cables or accessories.
Do not touch the power supply, X-ray Detector, cable, connector, or
any other electrical component or equipment with wet hands. Ignoring
this warning may cause electrical shock, which may result in severe
personal injury, death, or substantial product damage.
WARNING
Disconnect the cables by pulling on the connector and not the cable
itself. Ignoring this warning may cause electrical shock, which may
result in severe personal injury, death, or substantial product damage.
WARNING
Do not modify the cables or subject the cable to external stress or
damage. Avoid placing anything heavy, including the detector, on the
cable, stepping on the cable, pulling the cable, or subjecting the cable
to excessive bending or bundling. Ignoring this warning may cause
cable failure resulting in electrical shock, which may result in severe
personal injury, death, or substantial product damage.