www.vareximaging.com
57
1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
U
SER
M
ANUAL
13.4
RF Immunity
Table 27
Conducted/Radiated RF
Immunity test
IEC 60601-1-2
test level
Compliance
Electromagnetic environment
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
3 Vrms
150 kHz to 80
MHz
Portable and mobile RF communications
equipment should be used no closer to any part of
the detector than the recommended separation
distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended Separation Distance:
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7
GHz
10 V/m
80 MHz to 2.7
GHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
should be less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Some degradation of video & increase in video
noise possible.
d
3.5
V
1
-------
P
=
d
3.5
E
1
-------
P
=
d
7
E
1
------
P
=