Use of a carrier gas filter is required for optimum performance and to protect your system from
potential contamination. Your GC Accessory Kit includes a GC/MS Gas Clean Oxygen/Moisture
filter (CP-17973) and filter base. This easy-to-remove gas filter cartridge combines three highly
adsorptive materials in one filter to remove water, oxygen, and organic compounds to purify the
GC/MS gas. Carrier gas filters should be installed in a location where the indicator is visible.
The filter should be replaced when the indicator shows that the filter is saturated, or after one
year of service. For replacement procedures, refer to the instructions enclosed with your filter.
1200L LC/MS
Entrance
Before arranging for delivery of the 1200L LC/MS to your facility, please make sure that all
passages to the site of installation are at least 36 in. (91.5 cm) wide. Allow additional room for
maneuvering the shipping containers around corners and/or through doors.
CAUTION
The 1200L MS, foreline pumps, and Liquid Chromatograph
are heavy items that require at least two fit people for lifting
and moving safely.
Space and Load Requirements
The required layout for the 1200L MS with the LC is illustrated in Figure 3. The LC should be
positioned to the left of the instrument to minimize the delay volume between the LC and the
MS.
NOTE: This figure illustrates North American power requirements (115 Vac). See Table 3 for
230 Vac amperage requirements.
03-914961-00:6
10
of 24
1200/1200L MS Pre-installation Instructions