NOTE: This figure illustrates North American power requirements (115 Vac). See Table 4 for
230 Vac amperage requirements.
Figure 4 Recommended 1200L GC/LC/MS System Layout
The 1200L GC/LC/MS system should be installed on a clean, flat surface. The minimum bench
space required for the system is 57 cm x 250 cm for the HPLC, 1200, and PC, 325 cm with the
printer. Ensure that the bench is capable of supporting the weight of the system: at least 250 kg
plus allowances for any options.
Power Requirements 1200L GC/LC/MS
The 1200L GC/LC/MS system requires a dedicated duplex power source for the GC, two
dedicated fourplex single-phase power sources with earth grounds hard-wired to the main
power panel ground for the HPLC, MS, and computer, and one 230V 20A power source for the
foreline pumps. If you have additional sample preparation devices or test equipment, we
recommend a separate dedicated power source for their operation. Never plug the mass
spectrometer and the chromatograph into the same power source, or you may overload the
fourplex power source. Never use the free outlet on any of the power sources for equipment
drawing more than 2A
Within North America these power sources must be 20A, 120 Vac, 60 Hz
±
3 Hz and 20A,
230 Vac, 60 Hz
±
3 Hz for the two foreline pumps.
Outside North America power sources must be 10A, 230 Vac, 50 Hz
±
3 Hz and 20A, 230V
50 Hz
±
3 Hz for the two foreline pumps.
03-914961-00:6
14
of 24
1200/1200L MS Pre-installation Instructions