WARNING:
SHOCK HAZARD
Dangerous voltages are present. Unplug power cord.
When removing the filament screws do not allow ceramic dust to fall into the
ion trap. Use a flow of clean pressurized gas to blow off any dust observed.
Removing Old Filament Assemblies
1. Place the Source/Ion Trap Assembly on a lint-free surface with the filament
screws in the horizontal position. Do not stand the ion trap on its end or
ceramic dust may fall into the ion trap.
2. Remove the two Phillips screws.
3. Carefully lift the filament assembly out of the source holder.
4. Inspect the metal disc (03-931761-01)
on the magnet for discoloration and
carbon build up. If the disk looks dirty, slide the disc off and place a new one
on the magnet.
5. Turn the assembly over and repeat the steps for the other filament and
magnet disc.