Removing the Source/Ion Trap Assembly
1. Using your thumb and forefinger, gently wiggle each connector attached to
the external source or internal ionization assembly, while pulling the
connector off the pins. It is best to keep the internal source filament adaptor
attached to the flex cable if in internal ionization mode. Use tweezers or
pliers to get the connector off if necessary.
2. Loosen the two screws holding on the heat shield and slide it away from the
source or remove it.
3. Fully loosen the two screws that attach the trap assembly to the top flange.
4. Lift the assembly out and place the assembly on the holder provided (if
servicing the ion trap) or on a lint-free cloth with the source facing upwards.
Never rest the assembly on its heater connectors or source pins.
40 of 127