1. Align the EI volume as shown with large and small holes on opposite ends.
These holes need to align with the ion volume screw and transfer line hole.
2. Slide the EI ion volume into the source block.
3. Slide the CI ion volume into the source block so that the slot is aligned with
the ion volume screw.
4. Fully compress the spring. Ensure the proper hole in the CI ion volume aligns
with the transfer line hole.
5. Hold the CI ion volume in and slowly tighten the ion volume screw until it
stops. The CI ion volume should be captured by the screw entering the slot,
but still be able to freely slide in and out of the EI ion volume. Adjust the slot
position so that the CI ion volume slides freely after the screw is tightened.
Reinstalling the Lenses