4. Remove the filament retention screw.
5. Remove the ceramic plate.
6. Remove the filament assembly.
7. Remove the gate retaining screw.
8. The Center Ring is clipped around the end of the lens. Push a capillary pick
into the gap in the ring, lift and slide the ring over the edge of the gate.
9. Pull the insulator off the lens.
61 of 127