•
Thin blade knife (such as an X-acto® knife)
•
Pasteur
pipettes
•
Gloves - powder-free Nitrile, or lint-free cotton or lint-free nylon
•
Chemical wipes such as Kimwipes®
•
De-ionized
water
•
Isopropyl alcohol, methanol or methylene chloride
•
Acetone
•
Mild detergent (ph 6 to 7.5)
•
Aluminum
oxide
•
Cotton swabs
•
Sandpaper
Common Procedures
The procedures in this section are common to many of the maintenance
procedures described later in this section. The manual contains links to these
common procedures where appropriate. Click the link to jump to the common
procedure indicated, then press the back arrow located on the Navigation
Toolbar to return to the original procedure. The page numbers of these
procedures are also indicated in the manual. It may be necessary to use the
Acrobat View menu to display the Navigation toolbar, if it isn’t already visible.
Turning Off the Mass Spectrometer
WARNING:
BURN HAZARD
Allow heated zones to cool before disassembly.
1. Shut down the mass spectrometer through the Startup/Shutdown tab in
System Control. Click the Shut Down button in the upper left corner of the
screen. The heaters will be turned off and the speed of the turbo pump will
be gradually reduced to 35% of full speed. It may take several hours for full
shutdown and cooling to take place.
To speed the shutdown process, the system can be powered down and
purged with Nitrogen. See “Turning Off the Mass Spectrometer with Nitrogen
Purge” (page 36).
34 of 127
Summary of Contents for 4000 GC
Page 4: ......
Page 40: ...6 of 127...
Page 77: ...3 Check source connection pins for proper alignment and straighten as necessary 43 of 127...
Page 124: ...90 of 127...
Page 148: ...114 of 127...