4. Push the connectors onto the source pins. Each pin must be correctly
aligned to prevent the pins from being bent.
5. If a pin is not aligned, use a pair of tweezers to move the pin into alignment.
Reinstalling the Analyzer Assembly
Prior to installing the analyzer, check for any particles inside the manifold or on
the analyzer assembly. If necessary, blow out any particles with clean and
filtered compressed inert gas. Inspect the upper flange O-ring for particles and
clean if needed.