The controller to manifold cable (1)
•
The manifold lens cable (2)
•
The manifold power cable (3)
3. Gently push the transfer line assembly towards the manifold to check that it
slides all the way into the manifold and does not stop prematurely. If the
transfer line stops, remove the analyzer and check the tip on the transfer line. It
may be bent and need to be straightened or replaced to ensure proper
operation of the transfer line. Also, for external mode, be sure the hybrid mode
plug is not in place. Once the transfer line slides all the way in, turn it clockwise
to lock it into place.
Turning On the Mass Spectrometer
1. Make sure the vent in front of the mass spectrometer is closed (turned
clockwise completely).
46 of 127