Varian 900-LC Series Liquid Chromatograph
6.
Purge the pumps if new solvent is being used.
7.
Purge any air bubbles from the autosampler injection valve if
necessary.
8.
Turn on the column heater and allow the column(s) to come to
the correct temperature if necessary.
Caution
It is imperative that whenever the column heater is turned on, the
pump should be running with a non-absorbing solvent flowing
through the column to avoid column degradation.
9.
Turn the detector lamp(s) on.
Caution
It is imperative that whenever the Varian detector lamp is turned
on, the pump should be running with a non-absorbing solvent
flowing through the cell, with no air bubbles present.
For more information on getting your system prepared for use
please see the 900-LC Series HPLC Help.
1. Click the Windows®
Start
button.
2. Click
Programs.
3. Click
Galaxie™.
4. Click
Help.
5. Click
Varian Integrated Help.
6. Click the plus sign next to Getting Started.
7. Click
Preparing the LC System For Use
.
3.4
Configuring the Device Settings and System Parameters
The Device Settings allow you to program settings for each
component within your iHPLC as well as control flow rates, purge
the pumps, turn the detector lamps on, heat the column oven and
move the autosampler needle and fraction collector probe.
The System Parameters are settings that will download directly
Publication date: December 2007
47