Varian 900-LC Series Liquid Chromatograph
✒
Note
For
optimum analytical performance
, it is recommended that the
ambient temperature of the laboratory be between
20 and 25 °C
and
be held constant to within ±2 °C throughout the entire working day.
1.3.2
Electrical Power Supplies
All power supplies should be single phase AC, 3 wire system
(active, neutral, ground or two active and ground) and should be
terminated at an appropriate connection receptacle that is within
reach of the system power cable.
1.3.2.1
900-LC Series HPLC System Power Connections
Power supply phase
Single
Power supply, current rating and
overload protection
Between 3 and 5 A (for 220-240 V
countries)
Between 6 and 12 A (for 110-120 V
countries)
Plug supplied
Required wall socket
Australia
Complies with AS3112
To comply with AS3112
HPM 787 or Clipsal 2015 (250 V,
10 A)
USA
Complies with Nema 5-15P
To comply with Nema 5-15R
Hubbell 5262 (125 V, 15 A)
Canada
Complies with Nema 5-15P
Complies with CS 22.2 No.
42
To comply with Nema 5-15R
Hubbell 5262 (125 V, 15 A)
Europe
Complies with CEE 7 Sheet
VII
To comply with CEE 7 Sheet VII
(250 V, 10/16 A)
Table 2.
Power connections
1.3.2.2
Fuses
1FS1 & 1FS2
T10 A H250 V, 3AB SLO-BLO (for 220-240 V countries)
Publication date: December 2007
21