Varian 900-LC Series Liquid Chromatograph
in applications demanding high usage of corrosive materials, an
external cooling air supply system is provided. It is
strongly
recommended
that the cooling air be supplied from an
environmentally controlled area that is away from the instrument
exhaust and any other area where corrosive materials are stored
or used.
The cooling air system with flue, fan, ducting and supply cowl,
must provide a minimum positive flow of 6 cubic meters per minute
(200 cfm). The ducting should be corrosion-resistant and fireproof.
1.3.7
Weights and Dimensions
System unit
Width
Depth
Height
Weight
Varian 920-LC Analytical Liquid
Chromatograph
750 mm
30 in
725 mm
29 in
665 mm
27 in
133 kg
294 lb
Varian 940-LC Semi-Preparative Liquid
Chromatograph
750 mm
30 in
725 mm
29 in
665 mm
27 in
139 kg
307 lb
Varian 940-LC Preparative Liquid
Chromatograph
750 mm
30 in
725 mm
29 in
665 mm
27 in
137 kg
302 lb
Shipping dimensions for all LC systems
950 mm
38 in
895 mm
36 in
1165 mm
46 in
170 kg
375 lb
IBM® compatible (typical)
450 mm
18 in
770 mm
30 in
520 mm
20 in
(n/a)
Laser (typical)
500 mm
18 in
650 mm
30 in
200 mm
20 in
(n/a)
External Options
Fraction Collector
400 mm
16 in
600 mm
24 in
480 mm
19 in
18 kg
34 lb
Shipping dimensions
760 mm
30 in
500 mm
20 in
600 mm
24 in
31 kg
69 lb
Refractive Index Detector
296 mm
12 in
385 mm
15 in
212 mm
9 in
11 kg
25 lb
Shipping dimensions
775 mm
31 in
475 mm
19 in
460 mm
18 in
14 kg
31 lb
Evaporative Light Scattering Detector
200 mm
360 mm
415 mm
11 kg
Publication date: December 2007
25