Varian 900-LC Series Liquid Chromatograph
•
set the option to turn the lamps on as the detector(s) power up
•
select normalization for the PDA detector
•
synchronize with other modules can be selected to have the
system wait for when a contact closure is to be used to start
data acquisition
•
choose the tray type and cooling options in the autosampler
section
•
edit the manual injector options including the reset time,
analytical (SUM) loop size and preparative loop size
•
edit the reset time for a manual injector which is used to
switch the valve back into the idle position
•
choose which leak detector option will be performed when a
leak is detected
•
choose solvents to be used in the valve module section if you
have a Scale Up Module
•
with semi-prep and prep systems you can choose the pump
head sizes, sample delivery pump head size and pressure
units
•
in the fraction collector section you must choose your rack
type, tube volume and probe depth
•
save your settings
3.5
Creating a Method
To create a new method:
1. Click
File > New > Method
in the upper left corner of the
Galaxie™ software.
2. Choose the Project Name and System name from the drop-
down menus and then click
Next
.
3. Enter the Method Name and Description and then click
OK
.
✒
Note
Galaxie™ will open to the main method screen on the Integration
Events tab.
4. In the lower left corner click on
control
.
✒
Note
Galaxie™ will show the Device Select window. You should see a list of
your system components just to the left of the Device Select window.
50
Publication date: December 2007