background image

Using the UV-Vis detector Sync.doc, Page 1 of 2 

Using the UV-VIS Detector Sync Signals on the 920-LC/940-LC. 

 
UV-VIS detector Ready, Start, and Fault inputs can be triggered, and UV-VIS detector Ready, Start 
and fault outputs can be detected, using the signals from the Detector Sync connector at the rear of the 
Varian 920-LC/940-LC instrument. 
 

 
 
 
 
A circuit board with screw terminals to plug into the 
Detector Sync connector is available (P/N 02-102285-00) 
to be used with a customers own flying lead cable. This 
board does  not have any electronics on it, it just provides 
a terminal block to screw wires into. Alternatively  the 
customer can assemble their own cable with the desired 
terminations themselves. 
 
These circuit diagrams show the electrical interface of 
these signals inside the Varian instrument, to enable 
anyone to use these signals. 
 
 
 
 
 
 
 