1-1
959D Leak Detector
Section 1. Introduction
1.1 General Description
The Varian 959D is a portable, self-contained mass spectrometer helium leak detector.
Its basic components are:
❑
Test port that connects to the unit under test or to an associated vacuum system
❑
High vacuum pump
❑
Spectrometer tube
❑
Valve block
❑
TriScroll™ 610 roughing pump
❑
Leak detector electronics
❑
Operator interface
The unit provides clean and automatic leak detection utilizing a Contra-Flow™ vacuum sys-
tem design that allows testing at high test point pressures in a
dry
, oil-free vacuum pumping
system. The TriScroll dry roughing pump evacuates the test port to the appropriate transfer
pressure and provides ultra-clean system roughing and backing functions. A Turbo or
MacroTorr™ pump is used to achieve high vacuum in the spectrometer tube. The TriScroll
roughing pump is located on the cart below the 959D. An optional power probe accessory
allows the 959D to operate as a quick-response, high-sensitivity sniffing leak detector.
Operation is performed using primary controls located on the front of the 959D and second-
ary controls located behind a locking access door. The 959D front view is shown in
Figure 1-1. The leak detector electronics operate the spectrometer assembly, control the
high vacuum pump, control the valve block, and provide leak rate and system status infor-
mation to the operator through bar graph and LED indicators on the front panel, through
audio, or through an optional meter or chart recorder.
The 959D has nitrogen purge available. This feature eliminates a high helium background
which could interfere with testing. The use of this feature depends upon your test method
and test conditions. For example, using the helium spray test method in a poorly ventilated
area would likely require the use of the nitrogen purge. The 959D can detect leaks from as
large as 1x10
−
4
atm cc/sec to as small as 2x10
−
10
atm cc/sec (about 1 cc in 30 years). The
ultimate helium sensitivity is a factor of the configuration of the pumps.
Summary of Contents for 959D
Page 16: ...959D Leak Detector 1 2 D R A F T 9 1 3 0 2 Figure 1 1 Model 959D Leak Detector ...
Page 28: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 48: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 68: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 78: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 81: ......