959D Leak Detector
1-13
1.7 Specifications
Operating specifications for the 959D are listed in Table 1-3.
NOTE
For detailed specifications for the V70 or V70D high vacuum
pumps, TriScroll pump, or other optional equipment, refer to
their manuals.
Table 1-3
959D Operating Specifications
Parameter
Value
Response Time
Less than 2 seconds for helium, in accordance with AVS Standard 2.1
Amplifier Drift
Less than 2% of most sensitive scale in accordance with AVS Standard
2.1
Noise Level
Less than 2% of full scale, peak-to-peak, in accordance with AVS
Standard 2.1
Leak Indicator
50-segment bar graph with exponent display. Two sets of direct reading
sensitivity ranges comprise four decades in each range:
10
−
9
to 10
−
6
a
tm cc/sec (HI sensitivity, low speed)
10
−
8
to 10
−
5
atm cc/sec (LO sensitivity, high speed)
High Vacuum Pump
70 l/m, maintenance-free turbomolecular pump or MacroTorr pump
TriScroll roughing Pump
TriScroll 610 pump with 10 mTorr base pressure
Test Port
1-1/8 inch (28.5 mm) ID quick coupling, NW-25 adapter included
Operating Temperature
5
o
C to 35
o
C (41
o
F to 95
o
F)
Power Requirements
With TriScroll 610 pump:
❑
115 VAC, 50/60 Hz, 20 Amps
❑
220 VAC, 50/60 Hz, 10 Amps
Vacuum Gauges
Thermocouple gauges with ranges from ATM to 10
-3
Torr measure
pressure in the spectrometer tube and in the test port manifold to
monitor the test port pressures.
Summary of Contents for 959D
Page 16: ...959D Leak Detector 1 2 D R A F T 9 1 3 0 2 Figure 1 1 Model 959D Leak Detector ...
Page 28: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 48: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 68: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 78: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Page 81: ......